Procedure of PH Calibration of Technotrans


  1. Press and Hold the ? button for 5 sec (Till screen is changed).
  2. Press F2, then Press F5.
  3. Calibration screen opens at the unit control system.
  4. Remove, Clean and dry the PH probe.
  5. Immerse the PH probe into buffer solution PH 4.
  6. Press F4 to start the calibration, A PH value of 4.0 is displayed.
  7. Press the Enter key to confirm the value.
  8. Remove, Clean and dry the PH probe.
  9. Immerse the PH probe into buffer solution PH 7.
  10. Press F4 again, A PH value of 7.0 is displayed.
  11. Press the Enter key to confirm the value.
  12. Press F4 again to come out of calibration.
  13. Press F5- Calibration completed.


How to Latching the Encoder of Motor of Regioman Press


Latching Procedure:
        1. Switch Off M Drive power (24 V) from drive cabinet
        2. Disconnect Encoder connector from M Drive
        3. Hold roller from other side and remove Encoder
        4. Connect New Encoder and encoder connector
        5. Switch ON power
        6. Go to Baudis PC and restart sercos ring (MLC Section)
        7. Go to the Drive cabinet and Connect Calibration keypad
        8. Switch ON test mode for particular drive
        9. On Calibration keypad, orange light indicates no safe stop and green   light indicates safe stop
        10. Press button for Latch position (bottom button) - sound comes and button light starts flashing
        11. After some time press latch button again
        12. After some time light on latch button goes Off –it means latching finished
        13. Switch of the selector switch and remove the calibration panel

How to Reset the BnR Stacter PLC of Muller Martini


1 Switch off Main Switch.
2 Move the Position Switch 2 - (1 to 0 for stacker 1, 3 & others).
Move the Position Switch 2 - (2 to 0 for stacker 2, 4).
3 Switch On the Main Switch.
4 After Bootup again Switch off the Main Switch.
5 Move the Position Switch 2 - (0 to 1 for stacker 1, 3 & others).
Move the Position Switch 2 - (0 to 2 for stacker 2, 4).
6 Switch On the Main Switch.
7 PLC now Reseted.

How to Reprogram the Flash Card of Stacker Panel view of Muller Martini


1 Switch Off the Main Switch.
2 Remove the Flash Card into Stacter Panel View
3 Format the Flash Card with Fat system
4 Copy the Saved Program from Laptop to Flash Card.
5 Insert the Flash Card in Stacker Panel View.
6 Switch On the Main Switch.
7 Check the all Parameter
(Saved program path D:\Muller Martini\Stacker Panel View of Laptop)

How to Take Back up of Main PLC CP260 of Muller Martini


How to Take Back up of Main PLC CP260 of MM

  1. Connect Laptop to PLC using RS-232 cable.

  1. Open Download Tool Software that is saved in laptop.

  1. Click On Open Project option and select “NGR 3_13.dip” file and press OK.

  1. Path for NGR 3_13.dip file :- C:\Daten\MMVZ\DLT\CFG.

  1. Set Entry Point as “MainControl”.

  1. Set Scheme as “BC_MMVZ_Datamodule_upload_15.31-,38edit_V3”.

  1. Click on Set Scheme button and Select Data module upload option from first three options.

  1. Press “Set” button.

  1. Press Execute button.

  1. Computer will ask for disk 3/4.

  1. Insert floppy disk ¾ and Press OK .

  1. Back up started.

  1. All configurations of customer is saved in CONFIG folder in floppy disk ¾.

  1. After Back up finished take out the floppy disk.

  1. Back up finished.

Downloading Program to Main PLC CP260 of Muller Martini


Downloading Program to Main PLC CP260

  1. Connect Laptop to PLC using RS-232 cable.

  1. Open Download Tool Software that is saved in laptop.

  1. Click On Open Project option and select “NGR 3_13.dip” file and press OK.

  1. Path for NGR 3_13.dip file :- C:\Daten\MMVZ\DLT\CFG.

  1. Computer will ask for disk drive ¾.

  1. Insert floppy disk ¾ and Press OK and wait for next message.

  1. Computer will ask for disk drive 4/4.

  1. Insert floppy disk 4/4 and Press OK, wait till computer finishes copying data from floppy.

  1. Set Entry Point as “MainControl”.

  1. Set Scheme as “UserDefined”.

  1. Check Interface should be “Local”.

  1. Click on Set Scheme button.

  1. Select first three options i.e. ( first select and then press mouse right button).

@ Main Control

@ Delivery Unit Controller 1

@ Delivery Unit Controller 2

     Press “Set” button

  1. Press Execute button.

  1. After pressing Execute button Program downloading starts from computer to PLC. It may take 1 hour(approx.).

What is MLC



MLC- Multi Link Controller

• 1st slot is for Power Unit

• 2nd slot is for CPU

• 3rd slot is for Section-1 Control (Print Unit- 1 to 4)

• 4th slot is for Section-2 Control (Print Unit- 5 to 8)

• 5th slot is empty(here we can add more sections-maximum 6)