- Press
and Hold the ? button for 5 sec (Till screen is changed).
- Press
F2, then Press F5.
- Calibration
screen opens at the unit control system.
- Remove,
Clean and dry the PH probe.
- Immerse
the PH probe into buffer solution PH 4.
- Press
F4 to start the calibration, A PH value of 4.0 is displayed.
- Press
the Enter key to confirm the value.
- Remove,
Clean and dry the PH probe.
- Immerse
the PH probe into buffer solution PH 7.
- Press
F4 again, A PH value of 7.0 is displayed.
- Press
the Enter key to confirm the value.
- Press
F4 again to come out of calibration.
- Press
F5- Calibration completed.